Filter for separation of plasma and cellular blood components
The Haemoselect 0.5 / 0.7 plasma filter with a hollow fiber consisting of polyethersulfon is an integral part of the H.E.L.P. apheresis treatment unit, and is used for plasma separation in H.E.L.P. apheresis on the Plasmat Futura machine.
Material effective surface area: 0,45 m²
Pore size 0.5 / 0.7µm
Sieving coefficient of albumin
Immunoglobulines and lipoproteines: ~1
Filling volume blood side 50 ml
Plasma side 125 ml
Internal fiber diameter 300 µm
Inner wall thickness 100 µm
Maximum TMP 80 mmHg
Maximum blood flow 180 ml/min
Housing: polycarbonate
Potting polyurethane
The plasma filter is sterilized with ethyleneoxid and only for single use.